FDwfAnalogInTriggerConditionInfo(HDWF hdwf, int *pfstrigcond)
Parameters:
- hdwf Interface handle.
- pfstrigcond Variable to receive the supported trigger conditions.
The function above returns the supported trigger type options for the instrument. They are returned (by
reference) as a bit field. This bit field can be parsed using the IsBitSet Macro. Individual bits are defined using the
TRIGCOND constants in dwf.h. These trigger condition options are:
trigcondRisingPositive (This is the default setting):
o For edge and transition trigger on rising edge.
o For pulse trigger on positive pulse.
trigcondFallingNegative:
o For edge and transition trigger on falling edge.
o For pulse trigger on negative pulse.
FDwfAnalogInTriggerConditionSet(HDWF hdwf, TRIGCOND trigcond)
Parameters:
- hdwf Interface handle.
- trigcond Trigger condition to set.
The function above is used to set the trigger condition for the instrument.
FDwfAnalogInTriggerConditionGet(HDWF hdwf, TRIGCOND *ptrigcond)
Parameters:
- hdwf Interface handle.
- ptrigcond Variable to receive the current trigger condition.
The function above is used to set the trigger condition for the instrument.
FDwfAnalogInTriggerLevelInfo(
HDWF hdwf, double *pvoltsMin, double *pvoltsMax, int *pnSteps)
Parameters:
- hdwf Interface handle.
- pvoltsMin Variable to receive the minimum voltage level.
- pvoltsMax Variable to receive the maximum voltage level.
- pnSteps Variable to receive the number of voltage level steps.
The function above is used to retrieve the range of valid trigger voltage levels for the AnalogIn instrument in Volts.